China Factory for Denv-1 To 4 - Macro & Micro-Test Sample Release Reagent – Macro & Micro-Test
China Factory for Denv-1 To 4 - Macro & Micro-Test Sample Release Reagent – Macro & Micro-Test Detail:
Product name
Macro & Micro-Test Sample Release Reagent
Certificate
CE, FDA, NMPA
Features
Rapid: only 5 mins.
Easy: only 3 steps (1.add the sample 2.Mix upside and down 10 times 3. Leave it for 5 mins).
Main components
Name | Main components | Component specifications | Quantity |
Sample Release reagent | Dithiothreitol, sodium dodecyl sulfate (SDS), RNase inhibitor, surfactant, purified water | 0.5mL/Vial | 50 Vial |
Note: Components in different batches of kits are not interchangeable.
Storage conditions and shelf life
Store and transport at room temperature. The shelf life is 24 months.
Applicable instruments
Instruments and equipment during sample processing, such as pipettes, vortex mixers, water baths, etc.
Sample requirements
Freshly collected oropharyngeal swabs, nasopharyngeal swabs.
Precision
When this kit is used for extraction from the in-house precision reference CV for 10 replicates, the coefficient of variation (CV, %) of Ct value is not more than 10%.
Inter-batch difference
When the in-house precision reference is tested on three batches of kits under trial production upon repeated extraction and, the coefficient of variation (CV, %) of Ct value is not more than 10%.
Performance comparison
● Extraction efficiency comparsion
Efficiency comparsion of magnetic beads method and sample releaser |
||||
concentration |
magnetic beads method |
sample releaser |
||
orfab |
N |
orfab |
N |
|
20000 |
28.01 |
28.76 |
28.6 |
29.15 |
2000 |
31.53 |
31.9 |
32.35 |
32.37 |
500 |
33.8 |
34 |
35.25 |
35.9 |
200 |
35.25 |
35.9 |
35.83 |
35.96 |
100 |
36.99 |
37.7 |
38.13 |
undet |
The extraction efficiency of sample releaser was similar to that of magnetic beads method, and the concentration of pathogen could be 200Copies/mL.
● CV value comparison
Repeatability of sample releaser extraction |
||
concentration:5000Copies/mL |
ORF1ab |
N |
30.17 |
30.38 |
|
30.09 |
30.36 |
|
30.36 |
30.26 |
|
30.03 |
30.48 |
|
30.14 |
30.45 |
|
30.31 |
30.16 |
|
30.38 |
30.7 |
|
30.72 |
30.79 |
|
CV |
0.73% |
0.69% |
When tested at 5,000 copies /mL, the CV of orFab and N was 0.73% and 0.69%, respectively.
Product detail pictures:
Related Product Guide:
The organization upholds the philosophy of Be No.1 in good quality, be rooted on credit history and trustworthiness for growth, will keep on to provide previous and new customers from home and overseas whole-heatedly for China Factory for Denv-1 To 4 - Macro & Micro-Test Sample Release Reagent – Macro & Micro-Test , The product will supply to all over the world, such as: Serbia, Manila, UK, The development of our company not only needs the guarantee of quality, reasonable price and perfect service, but also relies on our customer's trust and support! In the future, we are going to continue with the most qualified and high quality service to offer the most competitive price, Together with our customers and achieve win-win! Welcome to inquiry and consult!
By Odelia from India - 2017.07.28 15:46
The goods we received and the sample sales staff display to us have the same quality, it is really a creditable manufacturer.
By Frank from Anguilla - 2018.06.30 17:29