Coxsackie Virus Type A16 Nucleic Acid
Product name
HWTS-EV025-Coxsackie Virus Type A16 Nucleic Acid Detection Kit (Enzymatic Probe Isothermal Amplification)
Certificate
CE
Epidemiology
This kit adopts Enzymatic Probe Isothermal Amplification (EPIA), and designs specific primers and RNA base-containing probes (rProbe) for the highly conserved region of Cox A16, and adds Bst enzyme and RNaseH enzyme at the same time, in which the left and right ends of the RNA base of rProbe are labeled with fluorescent groups and quencher, respectively. Use the DNA polymerase activity and strand displacement activity of Bst enzyme to amplify the target to be tested at a constant temperature, the RNaseH enzyme can cleave the RNA bases on the target-probe hybrid chain, so that the fluorescent group and quencher of rProbe are separated thereby fluorescing. In addition, the left fragment of the residual rProbe RNA base can be used as a primer to further extend to form a product, which further accumulates the product. The fluorescent signal is continuously accumulated with the formation of the product, thereby realizing the detection of the target nucleic acid.
Channel
FAM | Coxsackie virus type A16 |
ROX | internal contro |
Technical Parameters
Storage |
≤-18℃ |
Shelf-life |
12 months |
Specimen Type |
Freshly collected throat swabs |
CV |
≤10.0% |
Ct |
≤38 |
LoD |
2000Copies/mL |
Applicable Instruments |
Applied Biosystems 7500 Real-Time PCR SystemsSLAN-96P Real-Time PCR Systems
LightCycler®480 Real-Time PCR system Real-time Fluorescence Constant Temperature Detection System Easy Amp HWTS1600 |
Work Flow
Option 1
Recommended extraction reagent: Macro & Micro-Test Viral DNA/RNA Kit (HWTS-3001, HWTS-3004-32, HWTS-3004-48) and Macro & Micro-Test Automatic Nucleic Acid Extractor (HWTS-3006)
Option 2
Recommended extraction reagent: Macro & Micro-Test Sample Release Reagent (HWTS-3005-8).