Hantaan Virus Nucleic

Short Description:

This kit is used for the qualitative detection of hantavirus hantaan type nucleic acid in serum samples.


Product Detail

Product Tags

Product name

HWTS-FE005 Hantaan Virus Nucleic Acid Detection Kit (Fluorescence PCR)

Epidemiology

Hantavirus is a kind of enveloped, segmented, negative-strand RNA virus. Hantavirus is divided into two types: one causes Hantavirus pulmonary syndrome (HPS), and the other causes Hantavirus hemorrhagic fever with renal syndrome (HFRS). The former is mainly prevalent in Europe and the United States, and the latter is hemorrhagic fever with renal syndrome caused by Hantaan virus, which is common in China. The symptoms of hantavirus hantaan type mainly manifest as hemorrhagic fever with renal syndrome, which is characterized by high fever, hypotension, bleeding, oliguria or polyuria and other renal function impairment. It is pathogenic to humans and should be paid enough attention to. 

Channel

FAM hantavirus hantaan type
ROX

Internal Control

Technical Parameters

Storage

≤-18℃

Shelf-life 9 months
Specimen Type fresh serum
Ct ≤38
CV <5.0%
LoD 500 Copies/μL
Applicable Instruments Applied Biosystems 7500 Real-Time PCR System

Applied Biosystems 7500 Fast Real-Time PCR Systems

QuantStudio®5 Real-Time PCR Systems

SLAN-96P Real-Time PCR Systems

LightCycler®480 Real-Time PCR system

LineGene 9600 Plus Real-Time PCR Detection System

MA-6000 Real-Time Quantitative Thermal Cycler

BioRad CFX96 Real-Time PCR System

BioRad CFX Opus 96 Real-Time PCR System

Work Flow

Recommended extraction reagents: Macro & Micro-Test Viral DNA/RNA Kit (HWTS-3017) (which can be used with Macro & Micro-Test Automatic Nucleic Acid Extractor (HWTS-EQ011)) by Jiangsu Macro & Micro-Test Med-Tech Co., Ltd. The extraction should be conducted according to the IFU. The extraction sample volume is 200μL. The recommended elution volume is 80μL.

Recommended extraction reagents: Nucleic Acid Extraction or Purification Kit (YDP315-R). The extraction should be conducted according to the IFU. The extraction sample volume is 140μL. The recommended elution volume is 60μL.