Hepatitis B Virus RNA
Product name
HWTS-HP007 Hepatitis B Virus RNA Nucleic Acid Detection Kit (Fluorescence PCR)
Epidemiology
Quantitative detection of HBV-RNA levels in serum can better monitor the level of HBV cccDNA in hepatocytes, and at the same time avoid causing greater damage to patients. It is of great significance for the auxiliary diagnosis of hepatitis B virus (HBV) infection and the efficacy monitoring and drug withdrawal prediction of NAs antiviral therapy in CHB patients.
Channel
FAM | HBV-RNA |
ROX |
Internal Control |
Technical Parameters
Storage |
≤-18℃ |
Shelf-life | 9 months |
Specimen Type | serum |
Tt | ≤42 |
CV | ≤5.0% |
LoD | 100 copies/mL |
Specificity | The interference test results show that there is no effects on the determination of HBV RNA when the concentration of bilirubin in the serum is not more than 168.2μmol/mL, the concentration of hemoglobin produced by hemolysis is not more than 130g/L, and the concentration of blood lipids is not more than 65mmol/mL; there is no effects on the determination of HBV RNA when the total IgG concentration in serum is not more than 5mg/mL. The cross-reactivity test results show that there is no cross reactions between this kit and other viruses or bacteria detected by nucleic acid in blood samples (hepatitis C virus, human cytomegalovirus, epstein-barr virus, human immunodeficiency virus, hepatitis A virus, syphilis, human herpes virus type 6, herpes simplex virus type 1, herpes simplex virus type 2, influenza A virus, propionibacterium acnes, staphylococcus aureus, candida albicans) and the human genomes; after patients take lamivudine, telbivudine, adefovir dipivoxil and entecavir, when the viral load of hepatitis B in the body is above the minimum LoD of the kit, the kit can still be effectively detected. |
Applicable Instruments | Applied Biosystems 7500 Real-Time PCR SystemApplied Biosystems 7500 Fast Real-Time PCR Systems
SLAN-96P Real-Time PCR Systems (Hongshi Medical Technology Co., Ltd.) |
Work Flow
Option 1.
Viral DNA/RNA Isolation Kit (NA007-2) by GenMag Biotechnology Co., Ltd. The specific extraction steps please strictly follow the instruction manual.
Macro & Micro-Test General DNA/RNA Kit (HWTS-3017-50, HWTS-3017-32, HWTS-3017-48, HWTS-3017-96) (which can be used with Macro & Micro-Test Automatic Nucleic Acid Extractor (HWTS-3006C, HWTS-3006B)) by Jiangsu Macro & Micro-Test Med-Tech Co., Ltd. The extraction should be carried out according to the instruction manual, the extracted sample volume is 200μL, and the recommended elution volume is 80μL.