Mycoplasma Hominis Nucleic Acid
Product name
HWTS-UR004A-Mycoplasma Hominis Nucleic Acid Detection Kit (Fluorescence PCR)
Epidemiology
Sexually transmitted diseases (STDs) are still one of the important threats to global public health security, which can lead to infertility, premature fetal birth, tumorigenesis and various serious complications. Mycoplasma hominis exists in the genitourinary tract and can cause inflammatory responses in the genitourinary tract. The MH infection of the genitourinary tract can cause diseases like non-gonococcal urethritis, epididymitis, etc., and among females, which may cause inflammation of the reproductive system that spreads centered on the cervix. At the same time, the common complication of MH infection is salpingitis, and a small number of patients may have endometritis and pelvic inflammatory disease.
Channel
FAM | MH target |
VIC(HEX) | Internal Control |
Technical Parameters
Storage | Liquid: ≤-18℃ In dark |
Shelf-life | 12 months |
Specimen Type | urethral secretions, cervical secretions |
Ct | ≤38 |
CV | <5.0% |
LoD | 1000Copies/mL |
Specificity | There is no cross-reactivity with other STD infection pathogens, which are outside the detection range, and there is no cross-reactivity with chlamydia trachomatis, ureaplasma urealyticum, neisseria gonorrhoeae, mycoplasma genitalium, herpes simplex virus type 1, herpes simplex virus type 2, etc. |
Applicable Instruments | It can match the mainstream fluorescent PCR instruments on the market.
Applied Biosystems 7500 Fast Real-Time PCR Systems QuantStudio®5 Real-Time PCR Systems SLAN-96P Real-Time PCR Systems LightCycler®480 Real-Time PCR system LineGene 9600 Plus Real-Time PCR Detection System MA-6000 Real-Time Quantitative Thermal Cycler BioRad CFX96 Real-Time PCR System BioRad CFX Opus 96 Real-Time PCR System |
Work Flow
Option 1.
Recommended extraction reagent: Macro & Micro-Test Sample Release Reagent (HWTS-3005-8). The extraction should be conducted in strict accordance with the instructions.
Option 2.
Recommended extraction reagent: Macro & Micro-Test Viral DNA/RNA Kit (HWTS-3017-50, HWTS-3017-32, HWTS-3017-48, HWTS-3017-96) (which can be used with Macro & Micro-Test Automatic Nucleic Acid Extractor (HWTS-3006C, HWTS-3006B)) by Jiangsu Macro & Micro-Test Med-Tech Co., Ltd. The extraction should be strictly conducted according to the instructions. The recommended elution volume should be 80 μL.
Option 3.
Recommended extraction reagent: Nucleic Acid Extraction or Purification Reagent (YDP302) by Tiangen Biotech (Beijing) Co., Ltd.. The extraction should be conducted in strict accordance with the instructions and the recommended elution volume is 80µL.