Our advancement depends on the superior gear ,superb talents and consistently strengthened technology forces for Fetal Fibronectin Levels, H Pylori Infection, Combo Test Kit, Our merchandise are widely recognized and trusted by users and can meet up with continuously changing economic and social needs.
Short Lead Time for Vitamin Test At Home - Macro & Micro-Test Sample Release Reagent – Macro & Micro-Test Detail:

Product name

Macro & Micro-Test Sample Release Reagent

Certificate

CE, FDA, NMPA

Features

Rapid: only 5 mins.

Easy: only 3 steps (1.add the sample 2.Mix upside and down 10 times 3. Leave it for 5 mins).

Main components

Name Main components Component specifications Quantity
Sample Release reagent Dithiothreitol, sodium dodecyl sulfate (SDS), RNase inhibitor, surfactant, purified water 0.5mL/Vial 50 Vial

Note: Components in different batches of kits are not interchangeable.

Storage conditions and shelf life

Store and transport at room temperature. The shelf life is 24 months. 

Applicable instruments

Instruments and equipment during sample processing, such as pipettes, vortex mixers, water baths, etc. 

Sample requirements

Freshly collected oropharyngeal swabs, nasopharyngeal swabs.

Precision

When this kit is used for extraction from the in-house precision reference CV for 10 replicates, the coefficient of variation (CV, %) of Ct value is not more than 10%.

Inter-batch difference

When the in-house precision reference is tested on three batches of kits under trial production upon repeated extraction and, the coefficient of variation (CV, %) of Ct value is not more than 10%.

Performance comparison

● Extraction efficiency comparsion

Efficiency comparsion of magnetic beads method and sample releaser

concentration
copies/mL

magnetic beads method

sample releaser

orfab

N

orfab

N

20000

28.01

28.76

28.6

29.15

2000

31.53

31.9

32.35

32.37

500

33.8

34

35.25

35.9

200

35.25

35.9

35.83

35.96

100

36.99

37.7

38.13

undet

The extraction efficiency of sample releaser was similar to that of magnetic beads method, and the concentration of pathogen could be 200Copies/mL.

● CV value comparison

Repeatability of sample releaser extraction

concentration:5000Copies/mL

ORF1ab

N

30.17

30.38

30.09

30.36

30.36

30.26

30.03

30.48

30.14

30.45

30.31

30.16

30.38

30.7

30.72

30.79

CV

0.73%

0.69%

When tested at 5,000 copies /mL, the CV of orFab and N was 0.73% and 0.69%, respectively.

Macro & Micro-Test Sample Release Reagent10


Product detail pictures:

Short Lead Time for Vitamin Test At Home - Macro & Micro-Test Sample Release Reagent – Macro & Micro-Test detail pictures

Short Lead Time for Vitamin Test At Home - Macro & Micro-Test Sample Release Reagent – Macro & Micro-Test detail pictures

Short Lead Time for Vitamin Test At Home - Macro & Micro-Test Sample Release Reagent – Macro & Micro-Test detail pictures

Short Lead Time for Vitamin Test At Home - Macro & Micro-Test Sample Release Reagent – Macro & Micro-Test detail pictures

Short Lead Time for Vitamin Test At Home - Macro & Micro-Test Sample Release Reagent – Macro & Micro-Test detail pictures


Related Product Guide:

Our products are broadly regarded and reliable by end users and can meet up with constantly transforming financial and social requires of Short Lead Time for Vitamin Test At Home - Macro & Micro-Test Sample Release Reagent – Macro & Micro-Test , The product will supply to all over the world, such as: Mexico, Singapore, Jakarta, Wide selection and fast delivery to suit your needs! Our philosophy: Good quality, great service, keep improving. We've been looking forward that more and more oversea friends join in our family for further development near the future!
  • The factory workers have rich industry knowledge and operational experience, we learned a lot in working with them,we are extremely grateful that we can encount a good company has excellent wokers.
    5 Stars By Arthur from Toronto - 2018.09.23 18:44
    The sales manager is very patient, we communicated about three days before we decided to cooperate, finally, we are very satisfied with this cooperation!
    5 Stars By Jack from Belize - 2018.12.05 13:53
    Write your message here and send it to us